Title: BIOSOLVE-IV-registry: Safety and performance of the Magmaris scaffold: 12-month outcomes of the first cohort of 1,075 patients
Authors: Verheye, Stefan
Wlodarczak, Adrian
Montorsi, Piero
Torzewski, Jan
Bennett, Johan
Haude, Michael
Starmer, Gregory 
Buck, Thomas
Wiemer, Marcus
Nuruddin, Amin A B
Yan, Bryan P-Y
Lee, Michael K-Y
Issue Date: 2021
Source: Verheye S, Wlodarczak A, Montorsi P, Torzewski J, Bennett J, Haude M, Starmer G, Buck T, Wiemer M, Nuruddin AAB, Yan BP, Lee MK. BIOSOLVE-IV-registry: Safety and performance of the Magmaris scaffold: 12-month outcomes of the first cohort of 1,075 patients. Catheter Cardiovasc Interv. 2021 Jul 1;98(1):E1-E8. doi: 10.1002/ccd.29260. Epub 2020 Sep 3. PMID: 32881396; PMCID: PMC8359313.
Journal: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
Abstract: We aimed to assess the safety and performance of the Magmaris sirolimus-eluting bioresorbable magnesium scaffold in a large patient population. Magmaris has shown good outcomes in small-sized controlled trials, but further data are needed to confirm its usability, safety, and performance. BIOSOLVE-IV is an international, single arm, multicenter registry including patients with a maximum of two single de novo lesions. Follow-up is scheduled up to 5 years; the primary outcome is target lesion failure (TLF) at 12 months. A total of 1,075 patients with 1,121 lesions were enrolled. Mean patient age was 61.3 ± 10.5 years and 19.2% (n = 206) presented with non-ST-elevation myocardial infarction (NSTEMI). Lesions were 3.2 ± 0.3 mm in diameter and 14.9 ± 4.2 mm long; 5.1% (n = 57) were bifurcation lesions. Device success was 97.3% (n = 1,129) and procedure success 98.9% (n = 1,063). The Kaplan-Meier estimate of TLF at 12 months was 4.3% [95% confidence interval, CI: 3.2, 5.7] consisting of 3.9% target lesion revascularizations, 0.2% cardiac death, and 1.1% target-vessel myocardial infarction. Definite/probable scaffold thrombosis occurred in five patients (0.5% [95% CI: 0.2, 1.1]), thereof four after early discontinuation of antiplatelet/anticoagulation therapy. BIOSOLVE-IV confirms the safety and performance of the Magmaris scaffold in a large population with excellent device and procedure success and a very good safety profile up to 12 months in a low-risk population.
Description: Cairns & Hinterland Hospital and Health Service (CHHHS) affiliated author: Gregory Starmer
DOI: 10.1002/ccd.29260
Keywords: bioresorbable scaffolds;coronary artery disease;magnesium;NSTEMI;sirolimus