Title: Accuracy of quantitative echocardiographic measures of right ventricular function as compared to cardiovascular magnetic resonance
Authors: Burstow, Darryl J.
Stedman, Kathy
Hamilton-Craig, Christian R.
Maxwell, Ryan
Anderson, Bonita
Scalia, Gregory M.
Stanton, Tony 
Chan, Jonathan
Yamada, Akira
Issue Date: 2016
Source: 06/2303/22/received |05/02/accepted | 12 , 2016, p. 38-44
Pages: 38-44
Journal: International Journal of Cardiology. Heart & Vasculature
Abstract: BACKGROUND: Many echocardiographic parameters have been proposed to evaluate right ventricular (RV) systolic function. We comprehensively assessed a wide range of quantitative echocardiographic parameters in a single cohort compared with same-day cardiovascular magnetic resonance (CMR). METHODS AND RESULTS: 92 subjects were examined prospectively: Group 1 consisted of 46 healthy controls (21 males, 33.4 ± 11.4 years), Group 2 consisted of 46 patients (20 males, 38.5 ± 18.9 years) undergoing RV functional assessment by CMR (1.5 T). Echocardiography was performed on the same day as CMR; fractional area change (RVFAC), myocardial performance index via spectral Doppler (RVMPI), RVMPI via Doppler tissue imaging (RVMPI-DTI), peak systolic myocardial velocity by DTI (RVSm), tricuspid annular plane systolic excursion (TAPSE), speckle tracking strain, and three dimensional right ventricular ejection fraction (3DE-RV). Linear regression, Bland–Altman and receiver-operator-characteristic (ROC) analyses were performed. At ROC analysis, the most predictive echocardiographic methods were; RVFAC (AUC = 0.892), RVMPI (AUC 0.785), TAPSE (AUC 0.849) and 3DE-RV (AUC 0.909). 3DE-RV appeared the most accurate compared to CMR, although underestimated true RV volumes. CONCLUSION: As compared to CMR; 3DE-RV, RVFAC, TAPSE and RVMPI were the most reliable predictors of RV function. These parameters can be recommended for clinical use.S2352-9067(16)30027-6[PII]28616541[pmid]
Int J Cardiol Heart Vasc
DOI: 10.1016/j.ijcha.2016.05.007
Resources: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5454157/
Type: Article