Title: Investigation and Resolution of Interference in the LC-QTOF-MS Detection of 4-MePPP.
Authors: Darren Allen
Warnholtz C
McWhinney BC 
Issue Date: 2021
Journal: Journal of analytical toxicology
Abstract: An interference resulting in the false-positive detection of the synthetic cathinone 4-MePPP in urine was suspected following the recent addition of 4-MePPP spectral data to an LC-QTOF-MS drug library. Although positive detection criteria were achieved, it was noted that all urine samples suspected of containing 4-MePPP also concurrently contained high levels of tramadol and its associated metabolites. Using QTOF-MS software elucidation tools, candidate compounds for the suspected interference were proposed. To provide further confidence in the identity of the interference, in silico fragmentation tools were used to match product ions generated in the analysis with product ions predicted from the theoretical fragmentation of candidate compounds. The ability of the suspected interference to subsequently produce the required product ions for spectral library identification of 4-MePPP was also tested. This information was used to provide a high preliminary confidence in the compound identity prior to purchase and subsequent confirmation with certified reference material. A co-eluting isobaric interference was identified and confirmed as an in-source fragment of the tramadol metabolite, N,N-bisdesmethyltramadol. Proposed resolutions for this interference are also described and subsequently validated by retrospective interrogation of previous cases of suspected interference.
Type: Journal Article
Appears in Sites:Queensland Health Publications

Show full item record

Page view(s)

28
checked on Apr 24, 2025

Google ScholarTM

Check


Items in DORA are protected by copyright, with all rights reserved, unless otherwise indicated.