Please use this identifier to cite or link to this item: https://dora.health.qld.gov.au/qldresearchjspui/handle/1/11479
Title: A Comparison of Peripherally Inserted Central Catheter Materials
Authors: Ullman, A. J. ;August, D.;Kleidon, T. M. ;Walker, R. M.;Marsh, N.;Bulmer, A. C.;Pearch, B.;Runnegar, N.;Leema, J.;Lee-Archer, P.;Biles, C.;Gibson, V.;Royle, R.;Southam, K.;Byrnes, J.;Chopra, V.;Coulthard, A.;Mollee, P.;Rickard, C. M.;Harris, P. N. A.;Ware, R. S.
Issue Date: 2025
Source: N Engl J Med, 2025 (392) 2 p.161-172
Pages: 161-172
Journal Title: N Engl J Med
Abstract: BACKGROUND: New catheter materials for peripherally inserted central catheters (PICCs) may reduce the risk of device failure due to infectious, thrombotic, and catheter occlusion events. However, data from randomized trials comparing these catheters are lacking. METHODS: We conducted a randomized, controlled, superiority trial in three Australian tertiary hospitals. Adults and children who were referred for PICC placement were assigned in a 1:1:1 ratio to receive a hydrophobic or chlorhexidine PICC or a standard polyurethane PICC and were followed for 8 weeks. The primary outcome was device failure, which was a composite of infectious (bloodstream or local) or noninfectious (thrombosis, breakage, or occlusion) complications. RESULTS: A total of 1098 participants underwent randomization; 365 were assigned to the hydrophobic group, 365 to the chlorhexidine group, and 368 to the standard-polyurethane group. Device failure occurred in 21 of 358 participants (5.9%) in the hydrophobic group, in 36 of 363 (9.9%) in the chlorhexidine group, and in 22 of 359 (6.1%) in the standard-polyurethane group (risk difference, hydrophobic vs. standard polyurethane, -0.2 percentage points [95% confidence interval {CI}, -3.7 to 3.2; P = 0.89]; and chlorhexidine vs. standard polyurethane, 3.8 percentage points [95% CI, -0.1 to 7.8; P = 0.06]). In the hydrophobic group as compared with the standard-polyurethane group, the odds ratio for device failure was 0.96 (95% CI, 0.51 to 1.78), and in the chlorhexidine group as compared with the standard-polyurethane group, the odds ratio was 1.71 (95% CI, 0.98 to 2.99). Complications from any cause during the period of PICC placement occurred in 77 participants (21.5%) in the hydrophobic group, in 140 (38.6%) in the chlorhexidine group, and in 78 (21.7%) in the standard-polyurethane group (odds ratio, hydrophobic vs. standard polyurethane, 0.99 [95% CI, 0.69 to 1.42]; and chlorhexidine vs. standard polyurethane, 2.35 [95% CI, 1.68 to 3.29]). No adverse events were attributable to the interventions. CONCLUSIONS: Among adults and children who were referred for PICC placement, the risk of device failure due to noninfectious or infectious complications was not lower with hydrophobic or chlorhexidine PICCs than with standard polyurethane PICCs. (Funded by the National Health and Medical Research Council of Australia; PICNIC Australian New Zealand Clinical Trials Registry number, ACTRN12619000022167.).
DOI: 10.1056/NEJMoa2406815
Appears in Sites:Children's Health Queensland Publications
Queensland Health Publications

Show full item record

Page view(s)

184
checked on Oct 15, 2026

Google ScholarTM

Check

Altmetric


Items in DORA are protected by copyright, with all rights reserved, unless otherwise indicated.